Cross-talk suppressed multi-anode MCP-PMT

Physics – Instrumentation and Detectors

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

14 pages, 14 figures

Scientific paper

10.1016/j.nima.2008.04.014

We have developed a 4-channel multi-anode MCP-PMT, SL10, which exhibits a performance of sigma_TTS ~ 30 ps for single photons with G ~ 10^6 and QE=20% under a magnetic field of B <= 1.5 T. The cross-talk among anodes has been extensively studied. We have taken two measures to suppress it: one is to configure the SL10 to an effectively independent 4 small pieces of MCP-PMT's by segmenting an electrode of the second MCP-layer; the other is to use a constant fractional discriminator. Remarkable improvement has been achieved.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Cross-talk suppressed multi-anode MCP-PMT does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Cross-talk suppressed multi-anode MCP-PMT, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Cross-talk suppressed multi-anode MCP-PMT will most certainly appreciate the feedback.