Aging (Materials), Ambient Temperature, Annual Variations, Atomic Clocks, Crystal Oscillators, Environmental Engineering, Frequency Standards, Temperature Control, Time, Environmental Control, Frequency Measurement, Precision, Pressure Measurement, Regression Analysis, Rubidium, Temperature Measurement, Time Measurement
Scientific paper
The measurement of environmental parameters and frequencies of rubidium and precision crystal oscillators in two time and frequency laboratories are reported. The effects of ambient fluctuations to oscillator frequencies were studied using multiple linear regression. Laboratory temperature seems to be the most significant factor even in air conditioned rooms. Additional thermally controlled cabinets were built to reduce temperature effects.
Kalliomaeki Kalevi
Mansten Tapio
No associations
LandOfFree
Control of environmental and clock parameters in Time and Frequency Standards Laboratory does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Control of environmental and clock parameters in Time and Frequency Standards Laboratory, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Control of environmental and clock parameters in Time and Frequency Standards Laboratory will most certainly appreciate the feedback.