A high vacuum gate valve for skimmed molecular beam experiments is described. It is designed with a very short extent of only 10 mm along the molecular beam axis to minimize the distance between the molecular beam source and the experiment to provide the maximum molecular flux to the experiment. At the same time it provides free space on both sides of the skimmer to not disturb the supersonic expansion in front of the skimmer, to give optical access to the full distance between beam source and skimmer, and to allow for placing electrostatic devices very close behind the skimmer. The gate valve allows to maintain high vacuum conditions (10^{-8} mbar) in the experimental chamber while the source chamber is brought up to atmospheric pressure for modifications or maintenance. The valve can be operated from outside the vacuum chamber while maintaining vacuum conditions in all chambers.
Haak Henrik
Küpper Jochen
Meijer Gerard
Wohlfart Kirstin
No associations
LandOfFree
Compact in-place gate valve for molecular beam experiments does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Compact in-place gate valve for molecular beam experiments, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Compact in-place gate valve for molecular beam experiments will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-512906