Physics – Atomic Physics
Scientific paper
2006-02-23
Review of Scientific Instruments 77 (04/2006) 043107
Physics
Atomic Physics
Submitted to Review of Scientific Instruments
Scientific paper
10.1063/1.2194485
The average pixel distance as well as the relative orientation of an array of 6 CCD detectors have been measured with accuracies of about 0.5 nm and 50 $\mu$rad, respectively. Such a precision satisfies the needs of modern crystal spectroscopy experiments in the field of exotic atoms and highly charged ions. Two different measurements have been performed by illuminating masks in front of the detector array by remote sources of radiation. In one case, an aluminum mask was irradiated with X-rays and in a second attempt, a nanometric quartz wafer was illuminated by a light bulb. Both methods gave consistent results with a smaller error for the optical method. In addition, the thermal expansion of the CCD detectors was characterized between -105 C and -40 C.
Bigot Eric-Olivier Le
David Christian
Gotta Detlev
Hennebach Maik
Indelicato Paul
No associations
LandOfFree
Characterization of a CCD array for Bragg spectroscopy does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Characterization of a CCD array for Bragg spectroscopy, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Characterization of a CCD array for Bragg spectroscopy will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-711791