Calibration system in dual-channel photometer.

Physics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

Photometers: Calibration

Scientific paper

An internal calibration system for a dual-channel photometer has been made. The light from a single artificial source is passed simultaneously to two photomultipliers by using fiber light guides. Switching from working mode of the photometer to calibration is made automatically. The time required for channel reduction is aproximately 3 min, the accuracy of the calibration coefficients is about 0.02%, which introduces an error of no more than 0.0002m in resulting data.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Calibration system in dual-channel photometer. does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Calibration system in dual-channel photometer., we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Calibration system in dual-channel photometer. will most certainly appreciate the feedback.