Physics – Instrumentation and Detectors
Scientific paper
2010-08-30
Nucl.Instrum.Meth.A623:954-959,2010
Physics
Instrumentation and Detectors
8 pages, 5 figures
Scientific paper
10.1016/j.nima.2010.08.105
A method for the precise measurement of drift times for the high-precision drift chambers used in the TWIST detector is described. It is based on the iterative correction of the space-time relationships by the time residuals of the track fit, resulting in a measurement of the effective drift times. The corrected drift time maps are parametrised individually for each chamber using spline functions. Biases introduced by the reconstruction itself are taken into account as well, making it necessary to apply the procedure to both data and simulation. The described calibration is shown to improve the reconstruction performance and to extend significantly the physics reach of the experiment.
Grossheim A.
Hu Jianbin
Olin Art
No associations
LandOfFree
Calibration of the TWIST high-precision drift chambers does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Calibration of the TWIST high-precision drift chambers, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Calibration of the TWIST high-precision drift chambers will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-179353