Calibration of the STAR Forward Time Projection Chamber with Krypton-83m

Physics – Nuclear Physics – Nuclear Experiment

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

Scientific paper

The principles of the calibration of a time projection chamber with radioactive Krypton-83 are explained. The calculation of gain correction factors and the methods of obtaining a precise energy calibration are illustrated. The properties and advantages of \nuc{83m}{Kr} are summarized and compared to other radioactive calibration sources. It was shown that the Krypton calibration for the STAR FTPC is feasible and recommendable although the pad geometry of the FTPC causes a considerable deterioration of the measured spectrum due to only partially detected charge clusters.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Calibration of the STAR Forward Time Projection Chamber with Krypton-83m does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Calibration of the STAR Forward Time Projection Chamber with Krypton-83m, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Calibration of the STAR Forward Time Projection Chamber with Krypton-83m will most certainly appreciate the feedback.