Physics – Instrumentation and Detectors
Scientific paper
2008-11-18
Physics
Instrumentation and Detectors
4 pages, 6 eps figures. Talk given at the 24th ICNTS, Bologna, Italy (1-5 Sept. 2008)
Scientific paper
We developed a new method for determining the bulk etch rate velocity based on both cone height and base diameter measurements of the etched tracks. This method is applied here for the calibration of CR39 and Makrofol nuclear track detectors exposed to 158 A GeV In^{49+} and Pb^{82+} ions, respectively. For CR39 the peaks corresponding to indium ions and their different fragments are well separated from Z/beta = 7 to 49: the detection threshold is at REL ~ 50 MeV cm^2 g^{-1}, corresponding to a nuclear fragment with Z/beta = 7. The calibration of Makrofol with Pb^{82+} ions has shown all peaks due to lead ions and their fragments from Z/beta ~ 51 to 83 (charge pickup). The detection threshold of Makrofol is at REL ~ 2700 MeV cm^2 g^{-1}, corresponding to a nuclear fragment with Z/beta = 51.
Togo V.
Traore' I.
No associations
LandOfFree
Bulk etch rate measurements and calibrations of CR39 and Makrofol nuclear track detectors does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Bulk etch rate measurements and calibrations of CR39 and Makrofol nuclear track detectors, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Bulk etch rate measurements and calibrations of CR39 and Makrofol nuclear track detectors will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-101040