In particle therapy, knowledge of the stopping-power ratios (STPRs) of the ion beam for air and water is necessary for accurate ionization chamber dosimetry. Earlier work has investigated the STPRs for pristine carbon ion beams, but here we expand the calculations to a range of ions (1 <= z <= 18) as well as spread out Bragg peaks (SOBPs) and provide a theoretical in-depth study with a special focus on the parameter regime relevant for particle therapy. The Monte Carlo transport code SHIELD-HIT is used to calculate complete particle-fluence spectra which are required for determining STPRs according to the recommendations of the International Atomic Energy Agency (IAEA). We confirm that the STPR depends primarily on the current energy of the ions rather than on their charge z or absolute position in the medium. However, STPRs for different sets of stopping-power data for water and air recommended by the International Commission on Radiation Units & Measurements (ICRU) are compared, including also the recently revised data for water, yielding deviations up to 2% in the plateau region. In comparison, the influence of the secondary particle spectra on the STPR is about two orders of magnitude smaller in the whole region up till the practical range. The gained insights enable us to propose an analytic approximation for the STPR for both pristine and SOBPs as a function of penetration depth, which parametrically depend only on the initial energy and the residual range of the ion, respectively.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Analytical expressions for stopping-power ratios relevant for accurate dosimetry in particle therapy does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Analytical expressions for stopping-power ratios relevant for accurate dosimetry in particle therapy, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Analytical expressions for stopping-power ratios relevant for accurate dosimetry in particle therapy will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-159342

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.