The Cassegrain location of the high resolution optical spectrograph (HROS) for the 8-meter Gemini telescope presents a difficult challenge in controlling mechanical flexure. This is especially the case for a high-resolution spectrography, which requires large and heavy optical components. In HROS, to achieve the required spectrum stability of 2.0 micrometers /hr, we developed a closed-loop monitoring and correction system capable of measuring the spectrograph flexure in as it happens and actively compensating for image motion.
Bingham Richard G.
Charalambous Andrew
D'Arrigo Paolo
Saber-Sheikh Kambiz
Savidge Trevor E.
No associations
LandOfFree
Active flexure compensation for the HROS spectrograph does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Active flexure compensation for the HROS spectrograph, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Active flexure compensation for the HROS spectrograph will most certainly appreciate the feedback.