Absolute calibration of the Pierre Auger Observatory fluorescence detectors uses a light source at the telescope aperture. The technique accounts for the ombined effects of all detector components in a single measurement. The calibrated 2.5 m diameter light source fills the aperture, providing uniform illumination to each pixel. The known flux from the light source and the response of the acquisition system give the required calibration for each pixel. In the lab, light source uniformity is studied using CCD images and the intensity is measured relative to NIST-calibrated photodiodes. Overall uncertainties are presently 12%, and are dominated by systematics.
Bauleo Pablo
Brack J.
for the AUGER Collaboration
Garrard L.
Harton J.
No associations
LandOfFree
Absolute Calibration of the Auger Fluorescence Detectors does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Absolute Calibration of the Auger Fluorescence Detectors, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Absolute Calibration of the Auger Fluorescence Detectors will most certainly appreciate the feedback.