A method for the calibration of a Silicon Position Sensitive Detector is
described. From the combination of the signals collected at the ends of the
resistive electrode, energy and angle of the hitting particle are reconstructed
with a procedure which straightforwardly accounts for different amplifier chain
gains and dc offset effects.
Ostashko V.
Romano Silvano
Tumino A.
No associations
LandOfFree
A working with a Silicon Position Sensitive Detector (dE%E calibration and isotopes separation) does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with A working with a Silicon Position Sensitive Detector (dE%E calibration and isotopes separation), we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and A working with a Silicon Position Sensitive Detector (dE%E calibration and isotopes separation) will most certainly appreciate the feedback.