Physics – Instrumentation and Detectors
Scientific paper
2010-03-10
Physics
Instrumentation and Detectors
4 pages with 8 figures. Submitted to Nucl. Instr. and Meth. A (Elsevier) for proceedings of VCI 2010.
Scientific paper
This article presents a detector system consisting of three components, a CMOS imaging array, a gaseous-detector structure with a Micromegas layout and a UV-photon sensitive CsI reflective photocathode. All three elements have been monolithically integrated using simple post-processing steps. The Micromegas structure and the CMOS imaging chip are not impacted by the CsI deposition. The detector operated reliably in He/isobutane mixtures and attained charge gains with single photons up to a level of 6 \cdot 10^4. The Timepix CMOS array permitted high resolution imaging of single UV-photons. The system has an MTF50 of 0.4 lp/pixel which corresponds to app. 7 lp/mm.
Bilevych Yevgen
Breskin Amos
Carballo Victor Blanco
Cortesi Marco
der Graaf Harry van
No associations
LandOfFree
A UV Sensitive Integrated Micromegas with Timepix Readout does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with A UV Sensitive Integrated Micromegas with Timepix Readout, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and A UV Sensitive Integrated Micromegas with Timepix Readout will most certainly appreciate the feedback.