ASTROPARTICLE, PARTICLE AND SPACE PHYSICS, DETECTORS AND MEDICAL PHYSICS APPLICATIONS . Proceedings of the 9th Conference . Held
Physics
Medical Physics
Scientific paper
The feasibility of a subminiature scintillation detector to be inserted in a catheter for lesion localization in nuclear medicine SPECT has been studied. Measurements on a simple laboratory setup have been performed and compared with Monte Carlo results. Further simulations, at 30keV and 140keV, concerning a configuration reproducing severe clinical conditions have shown poor lesion detectability. Several factors affecting the response have to be investigated to improve the capability of lesion localization characterizing such detector.
Alonge G.
Burgio N.
Ciavola C.
Iurlaro G.
Montani L.
No associations
LandOfFree
a Subminiature Scintillation Detector for Catheter Operation does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with a Subminiature Scintillation Detector for Catheter Operation, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and a Subminiature Scintillation Detector for Catheter Operation will most certainly appreciate the feedback.