Physics – Instrumentation and Detectors
Scientific paper
2010-02-14
Physics
Instrumentation and Detectors
20 pages, 5 figures
Scientific paper
10.1016/j.nima.2010.04.152
We report on the development and performance of a high-sensitivity purity-analysis technique for gaseous xenon. The gas is sampled at macroscopic pressure from the system of interest using a UHV leak valve. The xenon present in the sample is removed with a liquid-nitrogen cold trap, and the remaining impurities are observed with a standard vacuum mass-spectroscopy device. Using calibrated samples of xenon gas spiked with known levels of impurities, we find that the minimum detectable levels of N2, O2, and methane are 1 ppb, 160 ppt, and 60 ppt respectively. This represents an improvement of about a factor of 10,000 compared to measurements performed without a coldtrap.
Dobi A.
Hall Carter
Kaufman Larry
Langford T.
Leonard Douglas Sidney
No associations
LandOfFree
A simple high-sensitivity technique for purity analysis of xenon gas does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with A simple high-sensitivity technique for purity analysis of xenon gas, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and A simple high-sensitivity technique for purity analysis of xenon gas will most certainly appreciate the feedback.