Nuclear Instruments and Methods in Physics Research Section A, Volume 273, Issue 2-3, p. 706-710.
Physics
3
Scientific paper
The performance of the position sensitive detector for the ZEBRA low energy gamma-ray imaging telescope is described. The detector consists of 9 position sensitive Nal(TI) elements each 5.8 × 5.0 × 56.0 cm viewed at either end of the long axis by 2 in. photomultiplier tubes. The total active area is 2470 cm2 with an average positional resolution of 2.1 cm and energy resolution of 15% FWHM at 661.6 keV. The method of flight calibration is described together with the provision within the on-board electronics to correct for sources of error in the calculation of event energy loss and position. The results presented are obtained from the calibration phase of the ZEBRA telescope project.
Butler Reginald C.
Caroli Ezio
Chiappetti Lucio
Court A. J.
Dean Anthony J.
No associations
LandOfFree
A position sensitive detector using a NaI(Tl)/photomultiplier tube combination for the energy range 200 keV to 10 MeV does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with A position sensitive detector using a NaI(Tl)/photomultiplier tube combination for the energy range 200 keV to 10 MeV, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and A position sensitive detector using a NaI(Tl)/photomultiplier tube combination for the energy range 200 keV to 10 MeV will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1759480