Scientific paper
10.1016/j.nima.2008.09.034
A large area plastic scintillator detector array(~ 1 m x1m) has been set up for fast neutron spectroscopy at the BARC-TIFR Pelletron laboratory, Mumbai. The energy, time and position response has been measured for electrons using radioactive sources and for mono-energetic neutrons using the 7Li(p,n1)7Be*(0.429 MeV) reaction at proton energies between 6.3 and 19 MeV. A Monte Carlo simulation of the energy dependent efficiency of the array for neutron detection is in agreement with the 7Li(p,n1) measurements. The array has been used to measure the neutron spectrum, in the energy range of 4-12 MeV, in the reaction 12C+ 93Nb at E(12C)= 40 MeV. This is in reasonable agreement with a statistical model calculation.
Chakrabarty D. R.
Datar V. M.
Kujur R.
Kumar Suresh
Mirgule E. T.
No associations
LandOfFree
A large area plastic scintillator detector array for fast neutron measurements does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with A large area plastic scintillator detector array for fast neutron measurements, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and A large area plastic scintillator detector array for fast neutron measurements will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-141824