Physics – Instrumentation and Detectors
Scientific paper
2005-02-21
Physics
Instrumentation and Detectors
5 pages, 7 figures
Scientific paper
Time of flight of a minimum ionizing particle along a fixed base has been measured with a 100 ps accuracy by means of a Dielectric Resistive Plate Chamber (DRPC) with 4 x 0.3 mm gas gaps. DRPC timing characteristics have been studied with different applied voltages, discriminating thresholds and beam intensities. It may be stated that the time-of-flight resolution of gaseous detectors developed within the ALICE experiment has reached the level of the best known scintillation counters.
Akindinov A.
Golovin V.
Martemiyanov A.
Petrov Victor
Plotnikov V. V.
No associations
LandOfFree
100 ps time-of-flight resolution of Dielectric Resistive Plate Chamber does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with 100 ps time-of-flight resolution of Dielectric Resistive Plate Chamber, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and 100 ps time-of-flight resolution of Dielectric Resistive Plate Chamber will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-197859