Evaluation of EUV toroidal diffraction gratings for SOHO/Coronal Diagnostic Spectrometer

Physics – Optics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

3

Scientific paper

Verification of component specification and EUV optical performance for the diffraction gratings of the normal incidence EUV spectrograph of SOHO/CDS is discussed. This discussion includes: a comparison of experimentally obtained EUV spectral images of a point source with corresponding ray trace analysis, in-band Lyman alpha leak determination, verification of toroidal radii of curvature, verification of blaze direction by calculation and VUV measurement, and zero order efficiency at 124 nm.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Evaluation of EUV toroidal diffraction gratings for SOHO/Coronal Diagnostic Spectrometer does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Evaluation of EUV toroidal diffraction gratings for SOHO/Coronal Diagnostic Spectrometer, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Evaluation of EUV toroidal diffraction gratings for SOHO/Coronal Diagnostic Spectrometer will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-957250

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.