WFC3 TV3 Testing: IR Gain Results

Physics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

Hubble Space Telescope, Hst, Space Telescope Science Institute, Wide Field Camera 3, Wfc3

Scientific paper

Over the course of Thermal Vacuum 3 (TV3) testing during the spring of 2008, we used flat field data to calculate the gain of the IR4 (FPA165) flight detector. At the nominal flight gain setting of 2.5 e-/ADU, we measure gain values between 2.31 - 2.41 e-/ADU, after accounting for inter-pixel capacitance (IPC). This corresponds to 2.66 - 2.77 e-/ADU before the IPC correction.