Homogeneity for highly asymmetric fields has been studied for a Siemens PRIMUS linear accelerator. The flattening filter has a radius smaller than the primary collimator one, creating inhomogeneities that affect large fields in areas far from the collimator axis, and asymmetric fields with large offset. Profiles and absolute dose have been measured in fields with two jaws at maximal position (20 cm) and the other two at maximal overtravel (10 cm.), corresponding to 10 x 10 fields with extreme offset. Profiles have a remarkable gradient decreasing towards the beam edge, making these fields unsuitable for treatments. Results show that the design of the primary collimator and flattening filter assembly has direct consequences in homogeneity. This can have clinical consequences for treatments involving fields that include these inhomogeneous areas. Comparisons with the treatment planning system (Philips Pinnacle) calculations, that computes under the hypotheses of a uniformly flattened beam, result in severe discrepancies.
Cutanda F.
Vargas-Castrillon S. T.
No associations
LandOfFree
The effect of the primary collimator and flattening filter on asymmetric fields for a Siemens PRIMUS linear accelerator does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with The effect of the primary collimator and flattening filter on asymmetric fields for a Siemens PRIMUS linear accelerator, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and The effect of the primary collimator and flattening filter on asymmetric fields for a Siemens PRIMUS linear accelerator will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-27205