Long Term Stability in Cw-Crds Experiments

Physics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

Mini-Symposium: Cavity Enhanced Spectroscopy

Scientific paper

Allan variance can be used to characterize the slow drift of CW cavity ring-down spectroscopy (CW-CRDS) system, caused by the instability of experimental environments. By including the differential measurement of on- and off-peak decay rates in the system, the drifting of both decay rates can cancel out greatly. A sensitivity of 5.6 × 10^{-12} cm^{-1} during an optimum integration time of 15.4 minutes has been obtained with our CW-CRDS system. This sensitivity corresponds a 3σ methane detection limit of 0.3 parts per billion by volume (ppbv) in N_2 at 20 torr or 37 parts per trillion by volume (pptv) in N_2 at 760 torr measurement pressure, near 1.65 μm laser wavelength.