Physics
Scientific paper
Apr 2005
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2005wfc..rept...15h&link_type=abstract
Instrument Science Report WFC3 2005-15, 6 pages
Physics
Hubble Space Telescope, Hst, Space Telescope Science Institute, Wide Field Camera 3, Wfc3
Scientific paper
During WFC3's initial thermal vacuum testing campaign in September and October 2004, the readout noise levels for the IR channel were measured. Readnoise values from both full frame and subarray ramps were in agreement with readnoise values previously measured on FPA64 at the Detector Characterization Lab (DCL). Using data from the IR01 and IR05 tests in thermal vacuum testing, WFC3's IR channel readnoise is between 19.5 e- /pix and 22.5 e-/pix for correlated double sampling (CDS), regardless of array size. The CEI Spec for IR channel CDS readnoise is 15 e-/pix.
No associations
LandOfFree
Results of WFC3 Thermal Vacuum Testing: IR Channel Readnoise does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Results of WFC3 Thermal Vacuum Testing: IR Channel Readnoise, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Results of WFC3 Thermal Vacuum Testing: IR Channel Readnoise will most certainly appreciate the feedback.