The Echelle spectrograph FOCES,1 that was operated at the 2.2m Calar Alto telescope between 1995 and 2009 was moved to the laboratories of Munich University Observatories and is being as a test bed for a number of different stability issues related to high precision radial velocity spectroscopy. We utilize FOCES to study spectrograph stability, illumination stability and fiber transport stability. Results from temperature and pressure stabilization are presented with this paper. We will show, that we reach the requirements set by our model analysis approach presented in [2]. Peak to valley mid term stability of temperature and pressure is as good as 0.002K and 0.02hPa.
Bender Ralf
Brucalassi Anna
Grupp Frank
Holzwarth Ronald
Hopp Ulrich
No associations
LandOfFree
Pressure and temperature stabilization of an existing chelle spectrograph II does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Pressure and temperature stabilization of an existing chelle spectrograph II, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Pressure and temperature stabilization of an existing chelle spectrograph II will most certainly appreciate the feedback.