Cycle 17: UVIS Bowtie Monitor

Physics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

Scientific paper

Ground testing revealed an intermittent hysteresis type effect in the UVIS detector {both CCDs} at the level of 1%, lasting hours to days. Initially found via an unexpected bowtie-shaped feature in flatfield ratios, subsequent lab tests on similar e2v devices have since shown that it is also present as simply an overall offset across the entire CCD, i.e., a QE offset without any discernable pattern. These lab tests have further revealed that overexposing the detector to count levels several times full well fills the traps and effectively neutralizes the bowtie. Each visit in this proposal acquires a set of three 3x3 binned internal flatfields: the first unsaturated image will be used to detect any bowtie, the second, highly-exposed image will neutralize the bowtie if it is present, and the final image will allow for verification that the bowtie is gone.;

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Cycle 17: UVIS Bowtie Monitor does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Cycle 17: UVIS Bowtie Monitor, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Cycle 17: UVIS Bowtie Monitor will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-1320835

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.