Physics
Scientific paper
Oct 2000
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2000wfc..rept....5c&link_type=abstract
Instrument Science Report WFC3 2000-05, 10 pages
Physics
Hubble Space Telescope, Hst, Space Telescope Science Institute, Wide Field Camera 3, Wfc3
Scientific paper
We present the results of charge transfer efficiency (CTE) measurements of a Marconi CCD. Thedevice is a model CCD44, similar to the WFC3 flight device, a CCD43. The device is coated to improve UV response. We present the results before and after the device has been exposed toone year's equivalent on-orbit radiation. CTE is measured with three techniques: 55Fe,Extended Pixel Edge Response (EPER) and First Pixel Response (FPR) (post-rad only).All data were collected at the Detector Characterization Lab (DCL) at Goddard Space FlightCenter.
Cawley Laura
Hanley Christopher
No associations
LandOfFree
WFC3 Detector CharacterizationReport # 1: CCD44 Radiation Test Results does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with WFC3 Detector CharacterizationReport # 1: CCD44 Radiation Test Results, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and WFC3 Detector CharacterizationReport # 1: CCD44 Radiation Test Results will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1107925