We show how nowadays it is possible to achieve the goal of accuracy and fast computation response in radiotherapic dosimetry using MonteCarlo methods, together with a grid computing model. We present a complete, fully functional prototype system for brachytherapy, integrating a Geant4-based simulation, an AIDA-based dosimetric analysis, a web-based user interface, and distributed processing either on a local computing farm or on geographically spread nodes. Thanks to the object-oriented approach adopted for the architecture, the work presented can be easily extended to become a general purpose dosimetric system, capable to address all radiotherapic techniques.
Foppiano F.
Guatelli Susanna
Moscicki J.
Pia Maria Grazia
No associations
LandOfFree
Distributed Processing, Montecarlo and CT Interface for Medical Treatment Planning does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Distributed Processing, Montecarlo and CT Interface for Medical Treatment Planning, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Distributed Processing, Montecarlo and CT Interface for Medical Treatment Planning will most certainly appreciate the feedback.