Computer Science
Calibrating, Cooling, Detectors, Helium, Lasers, Photometers, Radiometers, Accuracy, Cryogenics, Errors
Scientific paper
Laser sources whose intensity is determined with a cryogenic electrical substitution radiometer are described. Detectors are then calibrated against this known flux, with an overall error of 0.028 percent (0.3 mmag). Ongoing research has produced laser intensity stabilizers with flicker and drift of less than 0.01 percent. Recently, the useful wavelength limit of these stabilizers have been extended to 1.65 microns by using a new modular technology and InGaAs detector systems. Data from Si photodiode calibration using the method of Zalewski and Geist are compared against an absolute cavity radiometer calibration as an internal check on the calibration system.
No associations
LandOfFree
Absolute photometric calibration of detectors to 0.3 mmag using amplitude-stabilized lasers and a helium-cooled absolute radiometer does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Absolute photometric calibration of detectors to 0.3 mmag using amplitude-stabilized lasers and a helium-cooled absolute radiometer, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Absolute photometric calibration of detectors to 0.3 mmag using amplitude-stabilized lasers and a helium-cooled absolute radiometer will most certainly appreciate the feedback.