Sep 1990
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=1990spie.1308..244s&link_type=abstract
IN: Infrared detectors and focal plane arrays; Proceedings of the Meeting, Orlando, FL, Apr. 18, 19, 1990 (A91-36726 15-35). Bel
Computer Science
Cryogenic Equipment, Electronic Equipment Tests, Infrared Detectors, Leakage, Calibrating, Infrared Imagery, Standards, Storage Stability
Scientific paper
The results of tests for leakage performed on ten IR sensor dewars are presented, and the design principles of the new testing devices are discussed. The ultrasensitive leak detector used for testing is compared to conventional detectors. The superfine leak calibrator consisting of a tracer gas supply, an aliquot volume, a pressure transducer, temperature gage, and valves was used to measure leak rates in the E-4 to E-12 std cc He/s range. The testing method is explained, including the gases used, the quadrupole mass analyzer, the reference leak calibration, and the temperature coefficient of the reference leak. The test results of the IR sensor dewars are shown: seven showed leak rates in the E-15 std cc He/s range, two had no detectable leaks, and one had a mid-range E-14 leak. The shelf lives of the dewars are calculated based on the results. The vacuum integrity of small IR sensor dewars can be reliably tested to the range of 1E-15 std cc He/s using the ultrasensitive leak detector and the superfine leak calibrator.
Bergquist Lyle E.
Sasaki Tito Y.
No associations
LandOfFree
Leak testing of IR sensor dewars to 1E-15 std He/s does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Leak testing of IR sensor dewars to 1E-15 std He/s, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Leak testing of IR sensor dewars to 1E-15 std He/s will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1558038