WFC3 underwent instrument-level testing under ambient conditions during late June and early July, 2004 at Goddard Space Flight Center (GSFC). Dark current and readnoise values of the UVIS channel have been measured. At a detector temperature of -70oC, the dark current and readout noise are both well within contract-end item (CEI) specifications. Dark current values range from 0.94 to 1.08 e-/pix/hr, and are comparable to preship values measured at Ball. Measurements of the readnoise reveal values of ~3 electrons, also comparable to pre-ship values and values measured during the mini-ambient testing performed earlier in the year.
Baggett Sylvia
Hilbert Bryan
No associations
LandOfFree
WFC3 UVIS Dark Current and Readnoise from Ambient Testing does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with WFC3 UVIS Dark Current and Readnoise from Ambient Testing, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and WFC3 UVIS Dark Current and Readnoise from Ambient Testing will most certainly appreciate the feedback.