Other
Scientific paper
Jan 2010
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2010wfc..rept....2v&link_type=abstract
Instrument Science Report WFC3 2010-02, 12 pages
Other
Hubble Space Telescope, Hst, Space Telescope Science Institute, Wide Field Camera 3, Wfc3
Scientific paper
Electronic crosstalk produces a signal in one or more amplifier quadrants in response to a stimulus in another quadrant. In the WFC3 IR camera, the crosstalk arises during the readout and appears as a negative mirror image of the source. Positioned symmetrically opposite the source about the dividing line between each of the coupled readout amplifier quadrants, the crosstalk appears at a lower level than the surrounding background, at the level of ~1e-06 that of the source signal. The level is low enough that it should not be an issue for most programs; dithering can help mitigate the effect. This ISR characterizes the position, intensity and shape of the crosstalk effect as seen during TV3 ground testing.
Baggett Sylvia
Viana Aline Carneiro
No associations
LandOfFree
WFC3 TV3 Testing: IR Crosstalk does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with WFC3 TV3 Testing: IR Crosstalk, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and WFC3 TV3 Testing: IR Crosstalk will most certainly appreciate the feedback.