The Advanced X-Ray Astrophysics Facility (AXAF) is a high spatial resolution X-ray observatory scheduled for launch in 1998. One of its two focal plane instruments is the High Resolution Camera (HRC). The HRC consists of two microchannel plate detectors with photocathodes of CsI and KBR, respectively. Current approaches to modeling the HRC detector efficiency have focused on the response of the photocathode over the energy range 0.1 to 10 keV. In this paper, we present recent laboratory measurements of the quantum efficiency of coated and uncoated microchannel plates as a function of energy and angle of incidence of the X-Ray. An empirical model is fit to the data, and the results are used to predict the efficiency of HRC at the focal plane.
Flanagan Kathryn A.
Kearns D.
Lessing Joshua
Murray Stephen S.
Roll Rachel K.
No associations
LandOfFree
Soft x-ray quantum efficiency of microchannel plates with CsI and KBr photocathodes does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Soft x-ray quantum efficiency of microchannel plates with CsI and KBr photocathodes, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Soft x-ray quantum efficiency of microchannel plates with CsI and KBr photocathodes will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1713370