Analytical considerations and test results bearing on the performance of a CsI(Tl) scintillating crystal coupled with a photodiode when used as a gamma-ray spectrometer are addressed. Laboratory test results on a number of CsI(Tl) bars with different sizes, diffusive coatings, and preamplifier designs are presented. A suitable event selection electronic logic design is shown which reduces the effect of noise on the count rate while retaining the desired energy threshold.
Caroli Ezio
Dean Anthony J.
Di Cocco Guido
Ferrandi G.
Grant Kerr
No associations
LandOfFree
Performance of position sensitive gamma ray detector unit based on CsI(Tl) bars with photodiode readout does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Performance of position sensitive gamma ray detector unit based on CsI(Tl) bars with photodiode readout, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Performance of position sensitive gamma ray detector unit based on CsI(Tl) bars with photodiode readout will most certainly appreciate the feedback.