Astronomy and Astrophysics – Astronomy
Scientific paper
Sep 2008
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2008basi...36...99n&link_type=abstract
Bulletin of the Astronomical Society of India, Vol. 36, No. 2, p. 99-110
Astronomy and Astrophysics
Astronomy
1
Instrumentation : Polarimeter
Scientific paper
Calibration and characterization of a dual-beam polarimeter installed at Kodaikanal Tower Telescope (KTT) for spectropolarimetric observations are presented in this paper. It was found that a slit width of 48 μ is optimum for the spectrograph setup at KTT and corresponding spectral resolution is 32.47 mÅ. It was demonstrated that the precision in polarization measurement can be achieved better than 0.1 % by increasing the exposure time. However, the polarimetric calibration accuracy is limited to 0.35 % for Stokes Q and U parameters and 0.2 % for Stokes V parameter, mainly due to the uncertainty in the retardance of the calibration retarder. A comparison of the magnetic field measurement between the Solar Optical Telescope (SOT) onboard Hinode and KTT for an active region was made and a good match was found after spatial smearing of Hinode/SOT measurements by 5.12''.
Devendran P.
Hariharan
Nagaraju K.
Ramesh K. B.
Rangarajan K. E.
No associations
LandOfFree
On the performance of a dual-beam polarimeter at Kodaikanal Tower Telescope does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with On the performance of a dual-beam polarimeter at Kodaikanal Tower Telescope, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and On the performance of a dual-beam polarimeter at Kodaikanal Tower Telescope will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1434474