Scientific paper
A cost-effective methodology for operating the 349-pixel Cerenkov light imaging camera of the imaging element of the TACTIC γ-ray telescope array at stable single-counts rate and safe anode current values has successfully been tried out despite variations in the light of the night sky experienced by these pixels from time to time. Important details of this unconventional scheme are discussed and reference is made to the methods evolved for in situ absolute and relative gain calibrations of the camera pixels, which are particularly required for this mode of operation.
Bhat C. L.
Bhatt Nilay
Dhar V. K.
Goyal H. C.
Kanda Akinobu
No associations
LandOfFree
On a single-counts rate stabilization scheme employed in the imaging camera of the TACTIC γ-ray telescope does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with On a single-counts rate stabilization scheme employed in the imaging camera of the TACTIC γ-ray telescope, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and On a single-counts rate stabilization scheme employed in the imaging camera of the TACTIC γ-ray telescope will most certainly appreciate the feedback.