We present a method to measure the relative spectral response of the Pierre Auger Observatory Fluorescence Detector. The calibration was done at wavelengths of 320, 337, 355, 380 and 405 nm using an end-to-end technique in which the response of all detector components are combined in a single measurement. A xenon flasher and notch-filters were used as the light source for the calibration device. The overall uncertainty is 5%.
Bauleo Pablo
Brack J. T.
for the Pierre Auger Collaboration
Harton J. L.
Knapik Robert
No associations
LandOfFree
Multi-wavelength Calibration Procedure for the Pierre Auger Observatory Fluorescence Detectors does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Multi-wavelength Calibration Procedure for the Pierre Auger Observatory Fluorescence Detectors, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Multi-wavelength Calibration Procedure for the Pierre Auger Observatory Fluorescence Detectors will most certainly appreciate the feedback.