Computer Science – Performance
Scientific paper
Jun 2004
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2004jtf....27...54l&link_type=abstract
Journal of Time and Frequency (ISSN 1001-1544), Vol. 27, No. 1, p. 54 - 63 (2004)
Computer Science
Performance
Cesium Clock, Internal Operating Parameters, Characteristic Of Clock
Scientific paper
Based on the monitoring data of the internal operating parameters of the six HP5071A cesium clocks at NTSC, which were imported in 1997, the variations of the internal operating parameters of the three cesium clocks (Cs08, Cs11 and Cs16) in the period from starting operating to the ending of the service life of the cesium beam tube are analysed and the changes of the clock performance, which took place around the ending of the service life of the cesium beam tube, are discussed. To monitor the internal operating parameters of the clocks is helpfull for better maintaining and using the clocks.
Liu Chun-Xia
Qi Yi
Wang Gai-Xia
No associations
LandOfFree
Monitoring and analyses of the internal operating parameters of cesium clock HP5071A does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Monitoring and analyses of the internal operating parameters of cesium clock HP5071A, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Monitoring and analyses of the internal operating parameters of cesium clock HP5071A will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-854831