Computer Science – Performance
Scientific paper
Jul 2008
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=2008spie.7018e..74s&link_type=abstract
Advanced Optical and Mechanical Technologies in Telescopes and Instrumentation. Edited by Atad-Ettedgui, Eli; Lemke, Dietrich.
Computer Science
Performance
Scientific paper
The CTU (Cryogenics Translation Unit) is a low range (+/-1 mm) high resolution (<50 nm) translation unit to be used at cryogenic temperature (20K). The unit is a multipurpose device capable of fine closed loop positioning. This device can be used as active element in IR Instrumentation for compensating thermo-elastic deformation moving optical elements or sensors. CTU motion system is based in thin flexures deformation to assure repeatability and moves in closed loop mode by means of a fine linear actuator and a calibrated non contact capacitive sensor. This paper describes main design features, how cryogenic testing of main requirements was carried out (including methodologies used for calibration and submicron verification), tested performances, and main lesson learned during the development.
Belenguer Tomás
Díez Lucía
Lecina María
Moral Andoni
Moreno Raso Javier
No associations
LandOfFree
Cryogenic high resolution translation unit (CTU) does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Cryogenic high resolution translation unit (CTU), we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Cryogenic high resolution translation unit (CTU) will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-794576