We describe the design and performance of a cryogenic Fourier transform spectrometer (Cryo-FTS) operating at a temperature of approximately 15 K. The instrument is based on a porch-swing scanning mirror design with active alignment stabilization using a fiber-optic coupled diode laser and voice-coil actuator mechanism. It has a KBr beamsplitter and has been integrated into an infrared radiometer containing a calibrated Si:As blocked impurity band (BIB) detector. Due to its low operating temperature, the spectrometer exhibits very small thermal background signal and low drift. Data from tests of basic spectrometer function, such as modulation efficiency, scan jitter, spectral range, and spectral resolution are presented. We also present results from measurements of faint point-like sources in a low background environment, including background, signal offset and gain, and spectral noise equivalent power, and discuss the possible use of the instrument for spectral characterization of ground-based infrared astronomy calibration sources. The Cryo-FTS is presently limited to wavelengths below 25 micrometers but can be in principle extended to longer wavelengths with changes in beamsplitter and detector.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Cryogenic Fourier transform infrared spectrometer from 4 to 20 micrometers does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Cryogenic Fourier transform infrared spectrometer from 4 to 20 micrometers, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Cryogenic Fourier transform infrared spectrometer from 4 to 20 micrometers will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-1386420

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.