The paper describes the cryogenic design of the 8 - 12 micrometers beamcombiner instrument MIDI for the VLTI. Because the instrument is very sensitive at the whole thermal wavelength region, all optics and mechanics inside the dewar have to be cooled to temperatures at 40 K. The main contribution to the heat load is the radiation. We propose to use one liquid Nitrogen cooled radiation shield and a second shield cooled by the first stage of a closed cycle cooler with a refrigeration capacity of 35 W at 50 K and 1.5 W at 4.2 K. The main problems by using a closed cycle cooler are the vibrations which can disturb the interferometric measurements. We measured the vibrations during cooler operation and checked different vibration isolation possibilities.
Graser Uwe
Laun Werner
Lebong Matthias
Ortlieb Norbert
Rohloff Ralf-Rainer
No associations
LandOfFree
Cryo design for the VLTI MIDI instrument does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Cryo design for the VLTI MIDI instrument, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Cryo design for the VLTI MIDI instrument will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1480693