Scientific paper
May 1997
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=1997adspr..19.1297t&link_type=abstract
Advances in Space Research, Volume 19, Issue 9, p. 1297-1306.
Other
2
Scientific paper
Operational Calibration method of visible and IR channels of GMS-5 VISSR is as follows. In the visible channel, albedo is computed by the following equation. A={(C-b_0)/b_1}^2/a-V_0/a, where C and V_0 are digital counts from VISSR and V_0 is bias voltage corresponding to 0 albedo respectively, other coefficients are determined from prelaunch test data. A principle of IR channel calibration depends upon a linear relation between incoming IR radiance and output voltage of IR channel. Taking the space as a cold source and a calibration shutter in VISSR as a high source, a calibration regression equation to convert output voltage of IR channel into temperature is determined. A little correction is necessary to compensate non-linear response characteristics of IR detector at low temperature. To cope with a malfunction of calibration shutter (not frequent occurrence) a method of calibration without a shutter information is developed. The monitoring of solar radiance for one year indicates a small degradation in visible channel detectors. The degradation among 4 detectors is not uniform.
Itaya H.
Kurihara Susumu
Tokuno Masami
Tsuchiya Ken'ichi
No associations
LandOfFree
Calibration of VISSR on board GMS-5 does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Calibration of VISSR on board GMS-5, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Calibration of VISSR on board GMS-5 will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-832859