Calibration of the Rossi X-ray Timing Explorer Proportional Counter Array

Astronomy and Astrophysics – Astrophysics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

77 pages, 34 figures. accepted by Ap J Supplement for April 2006 issue. This paper and additional calibration information are

Scientific paper

10.1086/500659

We present the calibration and background model for the Proportional Counter Array (PCA) aboard the Rossi X-ray Timing Explorer (RXTE). The energy calibration is systematics limited below 10 keV with deviations from a power-law fit to the Crab nebula plus pulsar less than 1%. Unmodelled variations in the instrument background amount to less than 2% of the observed background below 10 keV and less than 1% between 10 and 20 keV. Individual photon arrival times are accurate to 4.4 micro-seconds at all times during the mission and to 2.5 micro-seconds after 29 April 1997. The peak pointing direction of the five collimators is known to a precision of a few arc-seconds

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Calibration of the Rossi X-ray Timing Explorer Proportional Counter Array does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Calibration of the Rossi X-ray Timing Explorer Proportional Counter Array, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Calibration of the Rossi X-ray Timing Explorer Proportional Counter Array will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-587841

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.