10 pages, 5 figures, submitted to Classical and Quantum Gravity

Scientific paper

We present a frequency modulation technique for calibration of the displacement actuators of the LIGO 4-km-long interferometric gravitational-wave detectors. With the interferometer locked in a single-arm configuration, we modulate the frequency of the laser light, creating an effective length variation that we calibrate by measuring the amplitude of the frequency modulation. By simultaneously driving the voice coil actuators that control the length of the arm cavity, we calibrate the voice coil actuation coefficient with an estimated 1-sigma uncertainty of less than one percent. This technique enables a force-free, single-step actuator calibration using a displacement fiducial that is fundamentally different from those employed in other calibration methods.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Calibration of the LIGO displacement actuators via laser frequency modulation does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Calibration of the LIGO displacement actuators via laser frequency modulation, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Calibration of the LIGO displacement actuators via laser frequency modulation will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-211315

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.