Computer Science
Scientific paper
Dec 1992
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=1992spie.1734..302k&link_type=abstract
In: Gamma-ray detectors; Proceedings of the Meeting, San Diego, CA, July 21, 22, 1992 (A93-30101 11-35), p. 302-308.
Computer Science
Compressed Gas, Gamma Rays, Ionization Chambers, Radiation Detectors, Xenon, Cylindrical Chambers, High Pressure, Ion Recombination
Scientific paper
Energy resolution and pulse distribution of a compressed gaseous xenon cylindrical detector were calculated. The analytical calculation took into account gamma-ray energy, fluctuation of electron-ion pairs, electron distribution, recombination, and H excess. The calculation was performed for a xenon density less than 0.6 g/cm and H excess less than 2 percent.
Dmitrenko Valery V.
Komarov V. B.
Ulin Sergey E.
Uteshev Z. M.
No associations
LandOfFree
Calculation of characteristics of compressed gaseous xenon gamma-ray detectors does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Calculation of characteristics of compressed gaseous xenon gamma-ray detectors, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Calculation of characteristics of compressed gaseous xenon gamma-ray detectors will most certainly appreciate the feedback.