Absolute Calibration and Characterization of the Multiband Imaging Photometer for Spitzer. II. 70 micron Imaging

Astronomy and Astrophysics – Astrophysics

Scientific paper

Rate now

  [ 0.00 ] – not rated yet Voters 0   Comments 0

Details

19 pages, PASP, in press

Scientific paper

10.1086/522675

The absolute calibration and characterization of the Multiband Imaging Photometer for Spitzer (MIPS) 70 micron coarse- and fine-scale imaging modes are presented based on over 2.5 years of observations. Accurate photometry (especially for faint sources) requires two simple processing steps beyond the standard data reduction to remove long-term detector transients. Point spread function (PSF) fitting photometry is found to give more accurate flux densities than aperture photometry. Based on the PSF fitting photometry, the calibration factor shows no strong trend with flux density, background, spectral type, exposure time, or time since anneals. The coarse-scale calibration sample includes observations of stars with flux densities from 22 mJy to 17 Jy, on backgrounds from 4 to 26 MJy sr^-1, and with spectral types from B to M. The coarse-scale calibration is 702 +/- 35 MJy sr^-1 MIPS70^-1 (5% uncertainty) and is based on measurements of 66 stars. The instrumental units of the MIPS 70 micron coarse- and fine-scale imaging modes are called MIPS70 and MIPS70F, respectively. The photometric repeatability is calculated to be 4.5% from two stars measured during every MIPS campaign and includes variations on all time scales probed. The preliminary fine-scale calibration factor is 2894 +/- 294 MJy sr^-1 MIPS70F^-1 (10% uncertainty) based on 10 stars. The uncertainty in the coarse- and fine-scale calibration factors are dominated by the 4.5% photometric repeatability and the small sample size, respectively. The 5-sigma, 500 s sensitivity of the coarse-scale observations is 6-8 mJy. This work shows that the MIPS 70 micron array produces accurate, well calibrated photometry and validates the MIPS 70 micron operating strategy, especially the use of frequent stimulator flashes to track the changing responsivities of the Ge:Ga detectors.

No associations

LandOfFree

Say what you really think

Search LandOfFree.com for scientists and scientific papers. Rate them and share your experience with other people.

Rating

Absolute Calibration and Characterization of the Multiband Imaging Photometer for Spitzer. II. 70 micron Imaging does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.

If you have personal experience with Absolute Calibration and Characterization of the Multiband Imaging Photometer for Spitzer. II. 70 micron Imaging, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Absolute Calibration and Characterization of the Multiband Imaging Photometer for Spitzer. II. 70 micron Imaging will most certainly appreciate the feedback.

Rate now

     

Profile ID: LFWR-SCP-O-234151

  Search
All data on this website is collected from public sources. Our data reflects the most accurate information available at the time of publication.