Computer Science
Scientific paper
Jul 1991
adsabs.harvard.edu/cgi-bin/nph-data_query?bibcode=1991gecoa..55.2075s&link_type=abstract
Geochimica et Cosmochimica Acta, vol. 55, Issue 7, pp.2075-2078
Computer Science
3
Scientific paper
Inductively-coupled Plasma-Mass Spectrometry (ICP-MS) with quadrupole mass analysis provides a very fast and sensitive method for the determination of 230 Th in marine sediments. A detection limit of 0.40 DPM (36 femto-moles) was determined for the method with a precision of better than 3% (1 ). This is comparable to that of -counting methods, but samples can be analyzed in five minutes, compared to up to two weeks counting time for -spectrometry. Sample preparation is also significantly simplified, thus providing a relatively quick and easy method for the determination of "excess" 230 Th accumulation histories in deep-sea sediment cores.
François Roger
Shaw Timothy J.
No associations
LandOfFree
A fast and sensitive ICP-MS assay for the determination of 230 Th in marine sediments does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with A fast and sensitive ICP-MS assay for the determination of 230 Th in marine sediments, we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and A fast and sensitive ICP-MS assay for the determination of 230 Th in marine sediments will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-862248