Long Wavelength Spectrometer (Lws), Blocked Impurity Band (Bib) Detectors, Quantum Yield, Gain Dispersion, Detective Quantum Efficiency
Scientific paper
Keck’s Long Wavelength Spectrometer (LWS), is the facilityinstrument used for imaging and spectroscopy in the wavelengthrange of 3 28 μm at the Keck Observatory. LWS uses an 128 × 128 Si:As blocked impurity band (BIB) array manufactured by the Boeing Corporation. This paper discusses the method used for optimizing the detector’s operating parameters at a temperature of 8.5 K and bias voltage of 1.2 V. A process for characterizing detective quantum efficiency of BIB detectors is also presented.
No associations
LandOfFree
Characterization of the Si:As Blocked Impurity Band (BIB) Detector in Keck's Long Wavelength Spectrometer (LWS) does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Characterization of the Si:As Blocked Impurity Band (BIB) Detector in Keck's Long Wavelength Spectrometer (LWS), we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Characterization of the Si:As Blocked Impurity Band (BIB) Detector in Keck's Long Wavelength Spectrometer (LWS) will most certainly appreciate the feedback.