The authors describe the design and performances of the four imaging proportional counters for the flight model of the stellar X-ray polarimeter (SXRP) to be flown on the spectrum X-gamma satellite at the focus of the SODART X-ray telescope. The SXRP will measure the linear polarization from celestial sources in an energy band between 2 and 15 keV. The authors describe the design and the performance of the X-ray detectors with particular care to the design philosophy for a space-borne focal plane polarimeter. They demonstrate the use of a wedge and strip pattern as a cathode which provides the desired position resolution in the whole energy band with moderate power consumption and only three preamplifiers per detector. The design, manufacturing techniques, and the performance of the wedge and strip position readout are described. Measurement of the background produced by impact of high-energy protons on different species of the material used to build the detector are presented. Also, the effect of high ionization charge density produced by end-of-range high-Z particles has been measured and the results are discussed.
Costa Edgar
Dwyer Joseph
Ford Edward
Kaaret Phil
Nenonen Seppo
No associations
LandOfFree
Proportional counters for the stellar X-ray polarimeter with a wedge and strip cathode pattern readout system. does not yet have a rating. At this time, there are no reviews or comments for this scientific paper.
If you have personal experience with Proportional counters for the stellar X-ray polarimeter with a wedge and strip cathode pattern readout system., we encourage you to share that experience with our LandOfFree.com community. Your opinion is very important and Proportional counters for the stellar X-ray polarimeter with a wedge and strip cathode pattern readout system. will most certainly appreciate the feedback.
Profile ID: LFWR-SCP-O-1054447