Over the course of Thermal Vacuum 2 (TV2) testing during the summer of 2007, we obtained several datasets that allowed us to calculate the gain of the IR1 (FPA129) detector. At the nominal flight gain setting of 2.0 e-/ADU, we measure gain values between 2.12 - 2.16 e-/ADU, after accounting for inter-pixel capacitance (IPC). This corresponds to 2.41 - 2.45 e-/ADU before the IPC correction. For the other (unsupported) gain settings of 2.5, 3.0, and 4.0 e-/ADU, we calculate actual gain values of 2.78 - 2.88, 3.33 - 3.48, and 4.50 - 4.76 e-/ADU respectively, after IPC correction.